Construction and correct use and method of biopsy forceps


In recent years, endoscopy equipment and personnel continue to popularize, biopsy forceps is a direct operation of mucosal tissue endoscope accessories, need to meet the use of standards, but due to the absence of biopsy forceps incomplete reasons, the general effect is not obvious, the resulting hospital transfer infection escort more and more attention. Therefore, it is necessary to find a good sterile method for endoscope accessories to prevent cross-infection in the hospital. More and more attention is paid to the sterility of endoscope and its accessories. If disposable biopsy forceps are used, the possibility of cross-infection can be completely removed. Let's learn about the structure and correct use and method of biopsy forceps!

In recent years, endoscopy equipment and personnel continue to popularize, biopsy forceps is the direct operation of mucosal tissue endoscope accessories, need to meet the standard, but because not found biopsy forceps use incomplete reason, the general effect is not obvious, the resulting cross infection in the hospital more and more attention. Therefore, it is necessary to find a good sterile method for endoscope accessories to prevent cross-infection in the hospital. More and more attention is paid to the sterility of endoscope and its accessories. If disposable biopsy forceps are used, the possibility of cross-infection can be completely removed. Let's learn about the structure, correct use and method of biopsy forceps together!

4aa814eb-fd8a-4d96-b5fe-c4b181bb6e0a.jpg

  (1) Front end: The front end is composed of two cup-shaped pliers valves that can be opened and closed.

1. Pliers body: The biopsy pliers body is made of stainless steel threaded tube, and the metal wire for switching and pulling the pliers mouth is installed inside. Due to the special structure of the threaded tube, substances such as tissue blood are easy to enter the inside, but it is not easy to clean completely, and it cannot be cleaned completely. The biopsy forceps is inconvenient to operate, and the opening and closing are not smooth, and it cannot be opened.

2. Operating handle: the ring of the operating handle is used to cover the thumb, and the wide circular groove is used to place the index finger and middle finger. Under the operation of these three fingers, the force is transmitted to the pliers valve through the traction line to open and close.

  (2) the use of biopsy forceps: biopsy forceps in operation and maintenance will affect the use of the endoscope, so need to pay full attention

1. Pre-inspection: Confirm that the biopsy forceps is processed and used within a certain period of time before use. Before inserting the endoscope forceps channel, check the opening and closing of the forceps flap. The specific method is to knead the biopsy forceps itself into a large circle (the diameter of the circle is about 20cm), and then perform multiple opening and closing operations to observe whether the opening and closing of the forceps flap is smooth. If there is 1-2 discomfort, generally do not use this biopsy forceps. Next, it is necessary to check the closure of the biopsy forceps. Take the thin paper like letter paper and clamp it with the biopsy forceps. The thin paper will not fall off as qualified. It is necessary to observe again whether the two cups of the forceps flap are completely aligned. If there is a deviation, please stop using immediately. Otherwise, the pliers tube will be injured.

2. Precautions during operation: Before inserting the pliers tube, the mouth of the pliers flap should be closed, but don't forget to close it too hard, the traction line may stretch, which will affect the opening and closing of the pliers flap. When inserting the trachea, please enter along the opening direction of the trachea of the pliers and do not rub with the mouth of the trachea. If there is resistance when entering, please release the angle button and enter as naturally as possible. If it still fails, please take the endoscope out of the body for testing, or replace other biopsy forceps with a thinner type.

The above is the structure of the biopsy forceps and the correct use and method, in order to freely open and close the biopsy forceps valve, it is necessary to often add lubricating oil to the joint, but pay attention to the selection of paraffin oil and other types of oil, do not forget to use the silicone oil in the endoscope shell. After applying lubricating oil, please wipe the excess oil, only keep the lubrication in the joint, do not be exposed.