Medical foot switch (lower)
Release time:
2022-04-25
Source:
Medical foot switch design fully consider ergonomic factors, combined with the medical industry long-term high-intensity operation characteristics, the overall operation of the product is simple and intuitive, to avoid the occurrence of misoperation.
Medical foot switchThe design fully considers ergonomic factors, combined with the characteristics of long-term and high-intensity operations in the medical industry, the overall operation of the product is simple and intuitive, and the occurrence of misoperation is avoided to the greatest extent. Waterproof grade
0: No protection
1: water droplets into the shell without influence
2: When the shell is tilted to 15 degrees, water droplets dripping into the shell have no effect
3: Water or rain from the 60 degree corner to the shell has no effect
4: liquid from any direction spilled into the shell no harm effect
5: water rinse without any harm
6: Can be used in the cabin environment
7: can be in a short period of time resistance to immersion (1m)
8: long time immersion under certain pressure
Regulations on waterproof test in IPxx grade (IPX6,IPX7,IPX8)
IPx 6
Method name: Intense water spray test;
Test equipment: the inner diameter of the nozzle is 12.5mm
Test conditions: the distance between the test sample and the water nozzle is 2.5m ~ 3m, and the water flow rate is 100 L/min(6000 L/h)
Test time: calculated according to the surface area of the tested sample shell, 1 min per square meter (excluding the installation area) less than 3 min
IPx 7
Method Name: Short-time water immersion test
Test equipment and test conditions: immersion tank. The dimensions shall be such that the distance from the bottom of the sample to the water surface is at least 1m after the sample is placed in the immersion tank. The distance from the top of the sample to the water surface is at least 0.15m.
Test time: 30 min
IPx 8
Method name: continuous diving test;
Test equipment, test conditions and test time: agreed by the supply and demand (buyer and seller) parties, its severity should be higher than IPX 7